BASIC MACHINE FUNCTION AND DESCRIPTION
The NEW innovative "LF" leak testing machine is designed specifically for laboratory or off-line leak inspection of vials / ampoules / blisters / ffs strips etc. The inspector includes state-of-the-art measuring and control systems.
TESTER FRAME AND HOUSING
ELECTRICAL, MECHANICAL, AND PNEUMATIC COMPONENTS
LEAK INSPECTION CONTROL SYSTEM
INTELLIGENT AUTOMATIC PARAMETERS VERIFICATION (IAV)
This feature will commence automatically upon start up of the machine. The machine will perform three verification tests:
Phase A:Cycle 1 - Electrovalves Test
In this primary phase the electro valves installed are tested and checked that each one efficiently opens and closes.
Phase B: Cycle 2 - Heads Equalization
In this secondary phase the test chamber is subjected to a vacuum pressure. Thus warming up the total system and remove any moisture residue from the testing heads. The vacuum pressure is then released.
Phase C: Cycle 3 – Recording of Statistical Head Parameters
In this final testing phase the statistical parameters for the testing head will be recorded and verified that all values are within tolerance limits. These parameters will be the actual values that will be compared during the A.C.P. procedure.
AUTOMATIC CLEANING OF THE TEST CHAMBER - ACP
This feature automatically dries the test chamber (Pocket) from residual moisture left by a previous faulty container. After a Pocket rejects a container due to a leak, some residual moisture of the product might remain inside the Pocket. This will hinder the result of the next leak test that the same Pocket will perform. Therefore, an automatic drying process will take place on this particular testing head. Therefore this particular Pocket will perform an empty cycle during which a continuous vacuum dries it. While under vacuum, any residue will be evaporated and sucked, therefore drying the Pocket. The next cycle will then also be an empty cycle but this time an actual test is done to make sure that all the parameters of the Pocket are within the original characteristics as recorded in Phase C of the IAV. test. If residue is still detected then another cleaning cycle will be performed until it will be perfectly dried. When the Pocket is dry it will automatically return to leak test containers and will now allow the next container to be tested. An OK signal will be displayed on the screen.
AUTO TEST
This test verifies that the transducer on the testing head can detect the leakage that is guaranteed. This feature is activated prior to the testing cycle. During the test, the vacuum is applied until the testing pressure is achieved. A calibrated flow valve is then actuated and simulates a pressure in the testing chamber corresponding to the minimum leakage that is expected to be detected. The vial is then rejected.
DUAL TESTING SYSTEM – VACUUM / PRESSURE (OPTIONAL)
The system operates with Vacuum decay or Pressure decay technology depending what type of test would be preferred. With a touch of a button the testing system can be changed. A specific product can be saved with the specific testing technology and hence when choosing the product to be tested from the database, the relative testing method is automatically selected. Vacuum decay is chosen for water-based products. If liquid escapes from the container, this will evaporate under vacuum and hence will change our vacuum level inside the test chamber. As a result the machine will see this product as a leaker. Pressure decay is chosen for oil-based products. If the container leaks, the pressure inside the test chamber will force air inside the container thus reducing the pressure level inside the test chamber. As a result the machine will see this product as a leaker.
OPERATION / TEST CYCLE
Filled and sealed containers are manually loaded into the appropriate test chamber. On pressing the cycle start button the testing process commences. During the leak test process the test chamber is subjected to a vacuum. The pressure pattern in the chamber is analyzed in a specific time. The smallest leak would increase the pressure in the test chamber and compared to preset values. Whether the test result is good or bad, the operator interface shows the result clearly on the screen accordingly.
Blister packs are manually loaded into the appropriate test chamber. On pressing the cycle start button the testing process commences. During the leak test process the test chamber is subjected to a vacuum. The force sensor monitors precisely the swelling of each tablet cell. If there is a leak this force reduces and hence shows that a leak is present. Whether the test result is good or bad, the operator interface shows the result clearly on the screen accordingly.
SPECIFICATIONS:
EXTRA OPTIONS: